QUICKSET 1PC FLEX DRILL BIT 25
Report
- Report Number
- 1818910-2014-24152
- Event Type
- Malfunction
- Date Received
- July 24, 2014
- Date of Event
- July 22, 2014
- Report Date
- July 22, 2014
- Manufacturer
- DEPUY ORTHOPAEDICS, INC. 1818910
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
(B)(4). EXAMINATION OF THE RETURNED INSTRUMENT CONFIRMS A PORTION OF THE TIP BROKE OFF. A COMPLAINT DATABASE SEARCH HAS IDENTIFIED A TREND FOR THIS FAILURE WITHIN THE 2274 QUICKSET DRILL FAMILY. PREVIOUS INVESTIGATIONS HAVE DETERMINED THAT USE ERROR IS THE LIKELY ROOT CAUSE. AS A RESULT OF TRENDING HEALTH HAZARD EVALUATION, DVA-106292-HHE WAS CONDUCTED. THE INVESTIGATION DID NOT ESTABLISH THE NEED FOR CORRECTIVE ACTION. NO EVIDENCE WAS FOUND OF PRODUCT OR DESIGN ERROR AS A CONTRIBUTING FACTOR. SUBSEQUENT COMPLAINTS WILL BE MONITORED UNDER POST MARKET SURVEILLANCE SCP 1405 / SEP 419. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY..
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
DRILL BIT BROKE.