OT VERIO IQ METER
Report
- Report Number
- 2939301-2014-18315
- Event Type
- Malfunction
- Date Received
- July 24, 2014
- Report Date
- July 14, 2014
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 2 ¿ (11/12/2014). THE PATIENT¿S TEST STRIPS HAVE BEEN RETURNED ON 8/8/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 10/28/2014 WITH THE FOLLOWING FINDINGS: ADDITIONAL TESTS WERE PERFORMED ON THE TEST STRIP VIAL AND THE DESICCANT TO CHECK THE STRUCTURAL INTEGRITY AND FOR POSSIBLE MOISTURE ISSUE. THE STRUCTURAL INTEGRITY OF THE TEST STRIP VIAL WAS FOUND TO BE INTACT DURING A GROSS LEAK TEST. THE DESICCANT WITHIN THE SUBJECT VIAL WAS FOUND TO BE SATURATED BY MOISTURE AND FAILED TGA TESTING. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
FOLLOW-UP # 1 ¿ (09/16/2014), THE PATIENT¿S TEST STRIPS HAVE BEEN RETURNED ON 8/8/2014 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 9/3/2014 WITH THE FOLLOWING FINDINGS: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT FAILED TESTING. THE TEST STRIPS WERE FOUND TO HAVE RESULTS ABOVE RANGE WHEN TESTED WITH CONTROL SOLUTION. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
ON (B)(6) 2014, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING INACCURATE HIGH RESULTS OF "129MG/DL, 134 MG/DL" AS COMPARED TO ANOTHER METER THAT READ "63MG/DL" WHEN TESTS WERE PERFORMED WITHIN 30 MINUTES. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED RESULTS DID NOT MEET LIFESCAN¿S ACCURACY CRITERIA AND THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 433621 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3608572 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |