HOMECHOICE PRO
Report
- Report Number
- 1416980-2014-23954
- Event Type
- Malfunction
- Date Received
- July 23, 2014
- Date of Event
- June 30, 2014
- Report Date
- June 30, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DC, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE HOMECHOICE (HC) DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB (PAL). THE EVENT HISTORY LOG REVIEW SHOWED NO KEYSTROKES, PROGRAMMING, OR USE RELATED EVENTS THAT INDICATED AND/OR CONTRIBUTED TO THE RITE (RETURNED INSTRUMENT TEST EVALUATION) FAILURE. THE EXTERNAL/INTERNAL INSPECTION WAS PERFORMED AND THE DEVICE PASSED, ALONG WITH ALL OF THE ELECTRICAL TESTING. DURING THE EVALUATION, THE FLUID WAS TRANSFERRED ABOVE THE RITE SPECIFIED LIMITS AND THE VOLUMETRIC ACCURACY TEST THAT FOLLOWED, SHOWED THAT THE DEVICE HAD FAILED. PAL PERFORMED A MORE DETAILED INSPECTION OF THE DOOR ASSEMBLY. THE INSPECTION SHOWED THAT THE PISTON FOAM WAS DETERIORATED AND THE DOOR PISTON WAS CRACKED. THE EVALUATION CONCLUDED THAT THE CAUSE OF THE FAILED FLUID VOLUME ACCURACY TESTING WAS A COMBINATION OF THE DETERIORATED PISTON FOAM AND A CRACKED DOOR PISTON. THE PISTON FOAM AND THE DOOR PISTON WERE REPLACED TO RESOLVE THE ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE (HC) DEVICE, A BAXTER TECHNICIAN DETERMINED THE HC MACHINE FAILED RITE (RETURNED INSTRUMENT TEST/EVALUATION) TESTING FOR FLUID VOLUME ACCURACY. DEVICE FAILED DURING EVALUATION, NO PATIENT INVOLVED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 430318 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |