FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 3954158 · Received July 23, 2014

Report

Report Number
1416980-2014-23954
Event Type
Malfunction
Date Received
July 23, 2014
Date of Event
June 30, 2014
Report Date
June 30, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE HOMECHOICE (HC) DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB (PAL). THE EVENT HISTORY LOG REVIEW SHOWED NO KEYSTROKES, PROGRAMMING, OR USE RELATED EVENTS THAT INDICATED AND/OR CONTRIBUTED TO THE RITE (RETURNED INSTRUMENT TEST EVALUATION) FAILURE. THE EXTERNAL/INTERNAL INSPECTION WAS PERFORMED AND THE DEVICE PASSED, ALONG WITH ALL OF THE ELECTRICAL TESTING. DURING THE EVALUATION, THE FLUID WAS TRANSFERRED ABOVE THE RITE SPECIFIED LIMITS AND THE VOLUMETRIC ACCURACY TEST THAT FOLLOWED, SHOWED THAT THE DEVICE HAD FAILED. PAL PERFORMED A MORE DETAILED INSPECTION OF THE DOOR ASSEMBLY. THE INSPECTION SHOWED THAT THE PISTON FOAM WAS DETERIORATED AND THE DOOR PISTON WAS CRACKED. THE EVALUATION CONCLUDED THAT THE CAUSE OF THE FAILED FLUID VOLUME ACCURACY TESTING WAS A COMBINATION OF THE DETERIORATED PISTON FOAM AND A CRACKED DOOR PISTON. THE PISTON FOAM AND THE DOOR PISTON WERE REPLACED TO RESOLVE THE ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE (HC) DEVICE, A BAXTER TECHNICIAN DETERMINED THE HC MACHINE FAILED RITE (RETURNED INSTRUMENT TEST/EVALUATION) TESTING FOR FLUID VOLUME ACCURACY. DEVICE FAILED DURING EVALUATION, NO PATIENT INVOLVED. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
430318 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1