FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3954094 · Received July 23, 2014

Report

Report Number
2032227-2014-04670
Event Type
Malfunction
Date Received
July 23, 2014
Date of Event
June 18, 2014
Report Date
June 18, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

ALL BUTTONS FUNCTION PROPERLY. HOWEVER, MOISTURE DAMAGE WAS FOUND ON KEYPAD TRACES. NO BUTTON ERROR ALARM NOTED DURING TESTING. THE INSULIN PUMP RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP AND CRACKED INSULIN PUMP BELT CLIP SLOT.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT HE WAS PROGRAMMING THE INSULIN PUMP WHEN HE RECEIVED A BUTTON ERROR ALARM. CUSTOMER DOES NOT RECALL ANY SIGNIFICANT EVENTS LEADING TO THE ALARM OR IF A BUTTON WAS PRESSED FOR MORE THAN 3 MINUTES. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
430537 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LNAB

Patients

Seq Age Sex Outcome Treatment
1 71 YR