FDA Adverse Event Malfunction Summary report: N

XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM

MDR report key: 3954021 · Received July 23, 2014

Report

Report Number
2024168-2014-04738
Event Type
Malfunction
Date Received
July 23, 2014
Date of Event
June 20, 2014
Report Date
June 30, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
NIQ
PMA / PMN Number
P110019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE LEAK WAS ABLE TO BE CONFIRMED. BASED ON A VISUAL AND FUNCTIONAL INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. A QUERY/REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TREATING A MILDLY CALCIFIED LESION IN THE DIAGONAL CORONARY ARTERY. DURING PREPARATION, THE 2.25X23MM XIENCE XPEDITION WAS FOUND TO HAVE A HOLE IN THE LUMEN AND WAS PULLING IN AIR. THE DEVICE WAS NOT USED. A 2.25X18MM XIENCE XPEDITION WAS THEN USED, BUT IT FAILED TO CROSS DUE TO THE PATIENT'S ANATOMY. A 2.25X12MM XIENCE XPEDITION WAS THEN USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
431760 XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT SYSTEM NIQ AV-TEMECULA-CT 3101141

Patients

Seq Age Sex Outcome Treatment
1