XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Report
- Report Number
- 2024168-2014-04738
- Event Type
- Malfunction
- Date Received
- July 23, 2014
- Date of Event
- June 20, 2014
- Report Date
- June 30, 2014
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- NIQ
- PMA / PMN Number
- P110019
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE LEAK WAS ABLE TO BE CONFIRMED. BASED ON A VISUAL AND FUNCTIONAL INSPECTION OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. A QUERY/REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.
IT WAS REPORTED THAT THE PROCEDURE WAS TREATING A MILDLY CALCIFIED LESION IN THE DIAGONAL CORONARY ARTERY. DURING PREPARATION, THE 2.25X23MM XIENCE XPEDITION WAS FOUND TO HAVE A HOLE IN THE LUMEN AND WAS PULLING IN AIR. THE DEVICE WAS NOT USED. A 2.25X18MM XIENCE XPEDITION WAS THEN USED, BUT IT FAILED TO CROSS DUE TO THE PATIENT'S ANATOMY. A 2.25X12MM XIENCE XPEDITION WAS THEN USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 431760 | XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM | DRUG ELUTING CORONARY STENT SYSTEM | NIQ | AV-TEMECULA-CT | 3101141 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |