FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3954007 · Received July 23, 2014

Report

Report Number
3004209178-2014-88100
Event Type
Malfunction
Date Received
July 23, 2014
Date of Event
June 9, 2014
Report Date
June 23, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED THE DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME AND EXCESSIVE NO DELIVERY ALARM TESTS. THE INSULIN PUMP FUNCTIONED PROPERLY. THE BOLUS WIZARD FUNCTIONED PROPERLY PER BOLUS WIZARD SAMPLE IN THE USER GUIDE. THE INSULIN PUMP HAD MINOR SCRATCHES ON THE DISPLAY WINDOW, A CRACKED RESERVOIR TUBE LIP AND CRACKED BATTERY TUBE THREADS.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED HIGH BLOOD GLUCOSE READINGS. CUSTOMER'S CURRENT BLOOD GLUCOSE READING IS 300 MG/DL AND HAS TREATED WITH A MANUAL INJECTION. TROUBLESHOOTING WAS PERFORMED AND THE DRIVE SUPPORT CAP AND ALL SETTINGS APPEARED TO BE NORMAL. AN AIR BUBBLE WAS ALSO NOTED IN THE TUBING. CUSTOMER ALSO MENTIONED THAT INSULIN PUMP IS NOT DELIVERING INSULIN AND THAT THE CORRECTION BOLUS IS INCORRECT. IT WAS NOTED THAT THE INSULIN PUMP DID NOT MAKE CORRECT CALCULATIONS BASED ON THE INFORMATION ENTERED. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
432278 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 63 YR