FDA Adverse Event Malfunction Summary report: N

VENTED AUTOFEED CHAMBER

MDR report key: 3953664 · Received July 23, 2014

Report

Report Number
9611451-2014-00646
Event Type
Malfunction
Date Received
July 23, 2014
Date of Event
June 26, 2014
Report Date
June 26, 2014
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) WHERE IT WAS VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED CRAZING AROUND THE SIDE OF THE CHAMBER DOME CLOSE TO THE BASE. CRACKS WERE ALSO NOTED ON THE CHAMBER DOME. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT 140410. CONCLUSION: BASED ON THE INSPECTION CARRIED OUT IT IS LIKELY THAT THE DAMAGE WAS CAUSED BY THE CHAMBERS COMING INTO CONTACT WITH A SOLUTION CONTAINING ETHANOL/ALCOHOL RESULTING IN ENVIRONMENTAL STRESS CRACKING OF THE CHAMBER DOMES. EVERY MR290 CHAMBER IS PRESSURE TESTED FOLLOWING THE MANUFACTURING PROCESS TO CHECK FOR ANY LEAKS PRESENT IN THE CHAMBER DOME DUE TO CRACKS AND OTHER CAUSES. ANY CHAMBER WHICH FAILS THIS TEST IS REJECTED. IN ADDITION, THE PRESSURE TEST IS FOLLOWED BY A VISUAL INSPECTION OF EACH CHAMBER. THIS SUGGESTS THE CHAMBER WAS DAMAGED AFTER IT WAS RELEASED FOR DISTRIBUTION. THE HOSPITAL STAFF ALSO REPORTED THAT THE CHAMBER STARTED TO LEAK AFTER EIGHT DAYS OF USE. OUR USER INSTRUCTIONS THAT ACCOMPANY THE MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER STATE THE FOLLOWING: DO NOT SOAK, WASH, STERILIZE, OR REUSE THIS PRODUCT. AVOID CONTACT WITH CHEMICALS, CLEANING AGENTS, OR HAND SANITIZERS. SET APPROPRIATE VENTILATOR ALARMS. PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT. MAXIMUM OPERATING PRESSURE: 8 KPA. (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) REPRESENTATIVE THAT WATER LEAKED FROM AN MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER AFTER EIGHT DAYS OF USE. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
431075 VENTED AUTOFEED CHAMBER AUTOFEED CHAMBER BTT FISHER & PAYKEL HEALTHCARE LTD MR290V 1404100304

Patients

Seq Age Sex Outcome Treatment
1 VERA VENTILATOR| FPH RT380 ADULT EVAQUA2 BREATHING CIRCUIT