FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3953611 · Received July 23, 2014

Report

Report Number
2531779-2014-21193
Event Type
Malfunction
Date Received
July 23, 2014
Report Date
July 15, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 DATE OF SUBMISSION 09/19/2014-PRODUCT ANALYSIS: THE DEVICE WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE COMPLAINT COULD NOT BE DUPLICATED WITH INVESTIGATION. REVIEW OF THE BLACK BOX REVEALED A RELATED ALARM. THE PUMP SUCCESSFULLY COMPLETED A PRIME SEQUENCE, BOLUS DELIVERIES, AND 24-HOUR EXERCISE TEST WITHOUT ISSUE OR ALARM. THE FORCE SENSOR CALIBRATION WAS OUT OF SPECIFICATION. NO DAMAGE, DEFECT, OR MOISTURE WAS OBSERVED TO THE PUMP¿S INTERNAL COMPONENTS. UNRELATED TO THE COMPLAINT, CORROSION WAS OBSERVED ON THE BATTERY CAP CONTACT. THE BATTERY COMPARTMENT WAS UNDAMAGED AND WITHOUT EVIDENCE OF MOISTURE INGRESS. NO LEAKS WERE FOUND DURING LEAK TESTING.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
430069 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 13 YR