FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3953443 · Received June 13, 2014

Report

Report Number
1720753-2014-05053
Event Type
Malfunction
Date Received
June 13, 2014
Date of Event
May 28, 2014
Report Date
June 13, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE CINE BRIDGE PCB AND CINE DRIVE WERE EVALUATED AND RESEATED. THE CINE DRIVE WAS REFORMATTED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A CINE DRIVE ERROR MESSAGE. THIS WOULD PREVENT THE CINE FUNCTION FROM OPERATING, CAUSING A LOSS OF SUBTRACTION, ROAD-MAPPING, OR OTHER CRITICAL VASCULAR CINE FUNCTIONS. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
349440 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1