FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3953443
·
Received June 13, 2014
Report
- Report Number
- 1720753-2014-05053
- Event Type
- Malfunction
- Date Received
- June 13, 2014
- Date of Event
- May 28, 2014
- Report Date
- June 13, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE CINE BRIDGE PCB AND CINE DRIVE WERE EVALUATED AND RESEATED. THE CINE DRIVE WAS REFORMATTED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED A CINE DRIVE ERROR MESSAGE. THIS WOULD PREVENT THE CINE FUNCTION FROM OPERATING, CAUSING A LOSS OF SUBTRACTION, ROAD-MAPPING, OR OTHER CRITICAL VASCULAR CINE FUNCTIONS. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.