FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3953418
·
Received June 13, 2014
Report
- Report Number
- 1720753-2014-05055
- Event Type
- Malfunction
- Date Received
- June 13, 2014
- Date of Event
- May 28, 2014
- Report Date
- June 13, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN INVESTIGATION. THE FFB WAS EVALUATED AND REPLACED. THE GIB AND BACKPLANE CABLES WERE EVALUATED AND RESEATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT BOOT UP. THIS RESULTED IN A LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 350664 | 9900 | FLUOROSCOPY X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |