FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3953406 · Received June 13, 2014

Report

Report Number
1720753-2014-05060
Event Type
Malfunction
Date Received
June 13, 2014
Date of Event
May 26, 2014
Report Date
June 13, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE FAILURE COULD NOT BE DUPLICATED. THE PS1 WAS SUSPECTED TO BE THE CAUSE AND WAS REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM INTERMITTENTLY WOULD NOT DISPLAY A FLUOROSCOPIC IMAGE. THIS RESULTED IN AN INTERMITTENT, RECOVERABLE LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.