FDA Adverse Event Malfunction Summary report: N

XMAX MOTOR

MDR report key: 3953246 · Received July 23, 2014

Report

Report Number
1045834-2014-12545
Event Type
Malfunction
Date Received
July 23, 2014
Date of Event
June 20, 2014
Report Date
June 24, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
ERL
PMA / PMN Number
PK965080
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE. THE DEVICE WAS TESTED AND IT WAS OBSERVED THAT THE NOSE PINS WERE LOOSE CAUSING THE ATTACHMENT NOT TO ATTACH. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO MISUSE AND/OR ABUSE. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNSPECIFIED NEURO SURGERY, IT WAS OBSERVED THAT THE MOTOR DEVICE COULD NOT ATTACH ATTACHMENTS. THERE WERE NO DELAYS TO THE PLANNED SURGICAL PROCEDURE AS A SPARE IDENTICAL DEVICE WAS AVAILABLE FOR USE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
431432 XMAX MOTOR DRILL, SURGICAL, ENT (PNEUMATIC) - HANDPIECE ERL DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1 ATTACHMENT DEVICE