XMAX MOTOR
Report
- Report Number
- 1045834-2014-12545
- Event Type
- Malfunction
- Date Received
- July 23, 2014
- Date of Event
- June 20, 2014
- Report Date
- June 24, 2014
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- ERL
- PMA / PMN Number
- PK965080
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE. THE DEVICE WAS TESTED AND IT WAS OBSERVED THAT THE NOSE PINS WERE LOOSE CAUSING THE ATTACHMENT NOT TO ATTACH. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO MISUSE AND/OR ABUSE. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
IT WAS REPORTED THAT DURING AN UNSPECIFIED NEURO SURGERY, IT WAS OBSERVED THAT THE MOTOR DEVICE COULD NOT ATTACH ATTACHMENTS. THERE WERE NO DELAYS TO THE PLANNED SURGICAL PROCEDURE AS A SPARE IDENTICAL DEVICE WAS AVAILABLE FOR USE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 431432 | XMAX MOTOR | DRILL, SURGICAL, ENT (PNEUMATIC) - HANDPIECE | ERL | DEPUY SYNTHES POWER TOOLS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ATTACHMENT DEVICE |