FDA Adverse Event Malfunction Summary report: N

ENDOSCOPIC MULTIPLE CLIP APPLIER

MDR report key: 39532 · Received August 30, 1996

Report

Report Number
1527736-1996-00047
Event Type
Malfunction
Date Received
August 30, 1996
Date of Event
August 19, 1996
Report Date
August 30, 1996
Manufacturer
ETHICON ENDO-SURGERY, INC.
Product Code
GDO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

H6: ADDED EVALUATION CODES (CODE 400=YIELDED JAWS). RESULTS OF EVALUATION: CONCLUSION: BASED UPON THE INQUIRY INFO RECEIVED AND THE VISUAL EXAMINATION, IT WAS CONCLUDED THAT THE JAWS HAD BECOME DAMAGED AND COULD NOT HOLD A CLIP PROPERLY, MAKING THE INSTRUMENT NON-FUNCTIONAL. NO CONCLUSION COULD BE REACHED AS TO HOW THIS DAMAGE OCCURRED. COMMENTS: IF THE JAWS ARE CLOSED OVER A HARD OBJECT, THE JAWS CAN BECOME DAMAGED AND WILL NOT HOLD A CLIP PROPERLY. EACH INSTRUMENT IS EVALUATED DURING THE ASSEMBLY PROCESS TO ENSURE THE JAWS HOLD CLIPS PROPERLY.

Description of Event or Problem · 1

THE DEVICE WAS USED DURING A LAPAROSCOPIC CHOLECYSTECTOMY. THE APPLIER WOULD NOT WORK PROPERLY. THE INSTRUMENT WAS SPITTING CLIPS THAT WERE UNFORMED. ANOTHER APPLIER WAS OPENED AND IT WORKED FINE. THERE WAS NO CONSEQUENCE TO THE PT. 08/23/1996 1825 IT WAS REPORTED 3 CLIPS FELL INTO THE PT AND ALL WERE RETRIEVED LAPAROSCOPICALLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOSCOPIC MULTIPLE CLIP APPLIER ENDOSCOPIC CLIP APPLIER GDO ETHICON ENDO-SURGERY, INC. NA J42M98

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other