HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2014-23873
- Event Type
- Malfunction
- Date Received
- July 23, 2014
- Date of Event
- June 28, 2014
- Report Date
- June 28, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER WAS UNKNOWN, A DEVICE ANALYSIS COULD NOT BE COMPLETED. HOWEVER, INITIATING THERAPY BEFORE CONNECTING IS A KNOWN CAUSE OF THIS ALARM. PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿ WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR WHEN AND HOW TO CONNECT TO THE DISPOSABLE SET. THE GUIDE ALSO INSTRUCTS TO CONNECT THE TRANSFER SET TO THE PATIENT LINE PRIOR TO STARTING THE INITIAL DRAIN. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A HOMECHOICE DEVICE PRESENTED A SYSTEM ERROR 2240 (AIR IN SET/LINE). THE PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. THIS OCCURRED DURING INITIAL DRAIN OF PERITONEAL DIALYSIS THERAPY. DURING TROUBLESHOOTING, THE PATIENT STATED THAT THEY HAD INITIATED THERAPY BEFORE CONNECTING TO THE PATIENT LINE. THE TECHNICAL SERVICE REPRESENTATIVE REVIEWED PROPER PROCEDURES WITH THE PATIENT. THE PATIENT PLANNED TO RESTART THERAPY USING NEW SUPPLIES. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 432210 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 89 YR | HOMECHOICE |