ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-21136
- Event Type
- Malfunction
- Date Received
- July 23, 2014
- Report Date
- July 13, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP # 1 DATE OF SUBMISSION 10/02/2014-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP BLACK BOX DATA REVEALED A VOLTAGE DROP THAT LED TO A CS 177 CALL SERVICE ALARM. NO OVERHEATING OR CALL SERVICE ALARMS WERE DUPLICATED DURING INVESTIGATION. CURRENT DRAWS MEASURED WITHIN SPECIFICATIONS. THE PUMP CASE WAS REMOVED AND NO ANOMALIES WERE FOUND IN THE POWER CIRCUIT.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A TEMPERATURE (TEMP - NO PHYSICAL DAMAGE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 432234 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 74 YR |