FDA Adverse Event Malfunction Summary report: N

DUAL TRIGGER ROTARY

MDR report key: 3952984 · Received July 23, 2014

Report

Report Number
0001811755-2014-02639
Event Type
Malfunction
Date Received
July 23, 2014
Date of Event
June 30, 2014
Report Date
June 30, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED.

Additional Manufacturer Narrative · 1

THE REPORTED RUN-ON WAS NOT ABLE TO BE DUPLICATED DURING DEVICE EVALUATION. HOWEVER, UPON DISASSEMBLY, THE E-BOX WAS FOUND TO BE LOOSE, WHICH CAN LEAD TO THE REPORTED EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING CLEANING AT THE USER FACILITY THE DEVICE RAN WITHOUT USER ACTIVATION AFTER THE TRIGGER WAS RELEASED. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING CLEANING AT THE USER FACILITY THE DEVICE RAN WITHOUT USER ACTIVATION AFTER THE TRIGGER WAS RELEASED. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
430323 DUAL TRIGGER ROTARY INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT HWE STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1