FDA Adverse Event Malfunction Summary report: N

PLUM XL PLS MIC/MAC

MDR report key: 3952744 · Received May 19, 2014

Report

Report Number
9615050-2014-03495
Event Type
Malfunction
Date Received
May 19, 2014
Date of Event
January 1, 2014
Report Date
April 23, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K952799
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AT THIS TIME THE CUSTOMER WILL NOT BE RETURNING THE DEVICE FOR EVALUATION. IF THE DEVICE IS RECEIVED, A FOLLOW-UP REPORT WILL BE SUBMITTED. THE DEVICE WAS NOT RETURNED TO HOSPIRA FOR TESTING AND INVESTIGATION; THEREFORE, ATTRIBUTION OF THE ISSUE TO THE DEVICE COULD NOT BE DETERMINED. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED A SWOLLEN AND OVERHEATED BATTERY. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT FOR A SWOLLEN AND OVERHEATED BATTERY. NO TRACKING INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPY. DURING TESTING AT THE USER FACILITY, IT WAS NOTED THAT THE BATTERY WAS SWOLLEN AND OVERHEATED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
295828 PLUM XL PLS MIC/MAC 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK