FDA Adverse Event
Malfunction
Summary report: N
2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLS
MDR report key: 3952607
·
Received May 15, 2014
Report
- Report Number
- 2937457-2014-00810
- Event Type
- Malfunction
- Date Received
- May 15, 2014
- Date of Event
- April 18, 2014
- Report Date
- April 18, 2014
- Manufacturer
- FRESENIUS MEDICAL CARE NORTH AMERICA
- Product Code
- KDI
- PMA / PMN Number
- K994267
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATIONS FINDINGS TO DATE INDICATED THE REPORTED MALFUNCTION OCCURRED DURING RECIRCULATION ADN PRIME (MACHINE SET-UP), AND NOT DURING DIALYSIS MODE. THE SUER VISUALLY OBSERVED THE SALINE BAG REFILLING WITH DIALYSATE DURING CIRCULATION. THERE HAVE BEEN NO ADVERSE EVENTS ASSOCIATED WITH THE REPORTED ISSUE. THE REPORT IS BEING INVESTIGATED BY THE MANUFACTURER VIA A CAPA. THE INVESTIGATION IS PENDING A SUPPLEMENTAL MDR WILL BE FILED AT THE COMPLETION OF THE INVESTIGATION.
Description of Event or Problem · 1
A USER FACILITY REPORTED A SALINE BAG BACK FILLED DURING RECIRCULATION MODE. A PATIENT WAS NOT CONNECTED TO THE MACHINE AT THE TIME OF THE INCIDENT. THERE WERE NO PARTS REPLACED ON THE DEVICE BY THE TECHNICIAN AND THE MACHINE WAS RETURNED TO SERVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 290841 | 2008K2 HEMODIALYSIS SYS. OLC/DIASAFE PLS | KDI | FRESENIUS MEDICAL CARE NORTH AMERICA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN SALINE (DISCARDED NOT-USED) |