FDA Adverse Event Malfunction Summary report: N

2008K2 HEMODIALYSIS SYS OLC/DIASAFE PLS

MDR report key: 3952586 · Received May 15, 2014

Report

Report Number
2937457-2014-00813
Event Type
Malfunction
Date Received
May 15, 2014
Date of Event
April 22, 2014
Report Date
April 22, 2014
Manufacturer
FRESENIUS MEDICAL CARE NORTH AMERICA
Product Code
KDI
PMA / PMN Number
K994267
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATIONS FINDINGS TO DATE INDICATED THAT THE REPORTED MALFUNCTION OCCURRED DURING RECIRCULATION AND PRIME (MACHINE SET-UP), AND NOT DURING DIALYSIS MODE. THE USER VISUALLY OBSERVED THE SALINE BAG REFILLING WITH DIALYSATE DURING CIRCULATION. THERE HAVE BEEN NO ADVERSE EVENTS ASSOCIATED WITH THE REPORTED ISSUE. THE REPORT IS BEING INVESTIGATED BY THE MANUFACTURER VIA A CAPA. THE INVESTIGATION IS PENDING A SUPPLEMENTAL MDR WILL BE FILED AT THE COMPLETION OF THE INVESTIGATION.

Description of Event or Problem · 1

A USER FACILITY REPORTED A SALINE BAG BACK FILLED DURING RECIRCULATION MODE. A PATIENT WAS NOT CONNECTED TO THE MACHINE AT THE TIME OF THE INCIDENT. THE MANUFACTURER REPLACED THE AIR SEPARATOR ASSEMBLY AND THE MACHINE WAS RETURNED TO SERVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
290839 2008K2 HEMODIALYSIS SYS OLC/DIASAFE PLS KDI FRESENIUS MEDICAL CARE NORTH AMERICA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN SALINE (DISCARDED - NOT USED)