FDA Adverse Event Malfunction Summary report: N

I-STAT CHEM8+ CARTRIDGE

MDR report key: 3952580 · Received July 23, 2014

Report

Report Number
2245578-2014-00033
Event Type
Malfunction
Date Received
July 23, 2014
Date of Event
June 24, 2014
Report Date
June 24, 2014
Manufacturer
ABBOTT POINT OF CARE
Product Code
JGS
PMA / PMN Number
K053110
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4) THE INVESTIGATION WAS COMPLETED ON 07/29/2014. RETURN PRODUCT WAS NOT AVAILABLE FOR INVESTIGATION. HOWEVER, RETAIN CARTRIDGES WERE TESTED AND ARE FUNCTIONING ACCORDING TO SPECIFICATION.

Description of Event or Problem · 1

ON (B)(6) 2014, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT POTASSIUM RESULT OF >9.0 ON A (B)(6) YEAR OLD FEMALE PATIENT SAMPLE. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. DATE: TIME: POTASSIUM: SAMPLE: (B)(6) 2014, 09:16, >9.0, A. (B)(6) 2014, 09:33, >9.0, B. THERE WERE NO INJURIES REPORTED WITH THIS EVENT. THE CUSTOMER STATES THAT THE SAMPLE WAS DRAWN FROM AN IV LINE. BASED ON THE INFORMATION AVAILABLE HEMOLYSIS COULD HAVE POTENTIALLY BEEN A FACTOR IN THE RESULT OF >9.0. THE INVESTIGATION IS UNDERWAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
431109 I-STAT CHEM8+ CARTRIDGE CHEM8+ CARTRIDGE JGS ABBOTT POINT OF CARE NA J14113

Patients

Seq Age Sex Outcome Treatment
1 58 YR