I-STAT CHEM8+ CARTRIDGE
Report
- Report Number
- 2245578-2014-00033
- Event Type
- Malfunction
- Date Received
- July 23, 2014
- Date of Event
- June 24, 2014
- Report Date
- June 24, 2014
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- JGS
- PMA / PMN Number
- K053110
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
(B)(4).
(B)(4) THE INVESTIGATION WAS COMPLETED ON 07/29/2014. RETURN PRODUCT WAS NOT AVAILABLE FOR INVESTIGATION. HOWEVER, RETAIN CARTRIDGES WERE TESTED AND ARE FUNCTIONING ACCORDING TO SPECIFICATION.
ON (B)(6) 2014, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT POTASSIUM RESULT OF >9.0 ON A (B)(6) YEAR OLD FEMALE PATIENT SAMPLE. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. DATE: TIME: POTASSIUM: SAMPLE: (B)(6) 2014, 09:16, >9.0, A. (B)(6) 2014, 09:33, >9.0, B. THERE WERE NO INJURIES REPORTED WITH THIS EVENT. THE CUSTOMER STATES THAT THE SAMPLE WAS DRAWN FROM AN IV LINE. BASED ON THE INFORMATION AVAILABLE HEMOLYSIS COULD HAVE POTENTIALLY BEEN A FACTOR IN THE RESULT OF >9.0. THE INVESTIGATION IS UNDERWAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 431109 | I-STAT CHEM8+ CARTRIDGE | CHEM8+ CARTRIDGE | JGS | ABBOTT POINT OF CARE | NA | J14113 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |