FDA Adverse Event Malfunction Summary report: N

VERSACARE BED

MDR report key: 3952461 · Received May 14, 2014

Report

Report Number
1824206-2014-01483
Event Type
Malfunction
Date Received
May 14, 2014
Date of Event
April 16, 2014
Report Date
April 16, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACCOUNT STATED THEY FOUND AFTER ZEROING THE BED IT IS READING A NEGATIVE WEIGHT. HILL-ROM TECHNICAL SUPPORT ASKED THE ACCOUNT TO RE-ZERO THE SCALE AND IF THE BED EXIT STILL DOES NOT SET THEY MAY NEED TO REPLACE THE SCALE POWER CONTROL BOARD. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. NO FURTHER INFORMATION IS AVAILABLE ON THE REPAIR OF THE BED AT THIS TIME. THE INVESTIGATION IS ONGOING, HOWEVER IF ANY ADDITIONAL RELEVANT INFORMATION IS IDENTIFIED FOLLOWING COMPLETION OF THE INVESTIGATION, THE ADDITIONAL RELEVANT INFORMATION WILL BE SUBMITTED IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

THE ACCOUNT REPORTED THE BED EXIT ALARM WILL NOT SET FROM EITHER SIDE RAIL. THE BED IS LOCATED IN THE BED SHOP AT THE FACILITY. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
288420 VERSACARE BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 3200

Patients

Seq Age Sex Outcome Treatment
1