COR/TRI POST STEM INSERT SHAFT
Report
- Report Number
- 1818910-2014-24054
- Event Type
- Malfunction
- Date Received
- July 23, 2014
- Date of Event
- July 17, 2014
- Manufacturer
- DEPUY ORTHOPAEDICS, INC. 1818910
- Product Code
- LXH
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
EXAMINATION OF THE RETURNED INSTRUMENT CONFIRMS THE COMPLAINT. PREVIOUS INVESTIGATIONS, INCLUDING EVALUATIONS BY A DEPUY MATERIAL SCIENTIST HAS DETERMINED THAT USE ERROR IS THE MOST LIKELY ROOT CAUSE. MANUFACTURING OR MATERIAL ERROR WAS NOT IDENTIFIED. IT IS HOWEVER RECOGNIZED THAT IMPROVEMENTS ARE POSSIBLE TO ALLEVIATE THIS ISSUE EVEN IF NOT USED PROPERLY. ECO 292374 WAS APPROVED AND COMPLETED ON JANUARY 15, 2010, WHICH INCLUDES IMPROVEMENTS TO BOLSTER THE DURABILITY OF THIS INSTRUMENT. HEALTH HAZARD EVALUATIONS PRA-860-2009-HHE IN MARCH 2009 AND DVA-105642-HHE IN JANUARY 2011 WERE HELD TO DETERMINE THE RISK INVOLVED WITH THE TIP OF A MODULAR STEM INSERTER BREAKING OFF DURING SURGERY. THE HHE FROM 2009 AND AGAIN IN 2011 CONCLUDED THAT NO FIELD ACTION WAS REQUIRED. THE COMPLAINT SAMPLE WAS MANUFACTURED IN JULY OF 2010; AFTER THE CHANGES. BECAUSE THERE HAVE NOW BEEN FAILURES REPORTED INVOLVING THE IMPROVED DESIGN, AN ADDITIONAL PRELIMINARY RISK ASSESSMENT, DVA-107766-PRA WAS INITIATED IN JANUARY 2013. THE PRELIMINARY RISK ASSESSMENT CONCLUDED THERE WAS NO ADDITIONAL OR NEW PATIENT HARM ASSOCIATED WITH THE DEVICES. ADDITIONAL CORRECTIVE ACTION IS NOT INDICATED AT THIS TIME. CONTINUE TO MONITOR THROUGH SEP 419 POST MARKET SURVEILLANCE. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
POSTERIOR CORAIL/TRI-LOCK INSERTER TIP BROKE DURING INSERTION OF FEMORAL STEM. DATE OF SURGERY: (B)(6) 2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 430692 | COR/TRI POST STEM INSERT SHAFT | HIP INSTRUMENT/TRIAL | LXH | DEPUY ORTHOPAEDICS, INC. 1818910 | MT0710 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR |