FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3952397 · Received July 23, 2014

Report

Report Number
2531779-2014-21072
Event Type
Malfunction
Date Received
July 23, 2014
Report Date
July 17, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 10/16/2014 ¿ DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/01/2014 WITH THE FOLLOWING FINDINGS: DURING TESTING, THE DISPLAY WAS FOUND TO BE DIM AND DISCOLORED. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED THAT THE BATTERY COMPARTMENT WAS CRACKED. ALSO UNRELATED TO THE COMPLAINT, EVALUATION REVEALED THAT THE RETURNED BATTERY CAP WAS DAMAGE/CRACKED. A TEST BATTERY CAP WAS USED TO COMPLETE ALL TESTING. IN ADDITION TO THE UNRELATED ISSUES, EVALUATION REVEALED THAT THE OK AND CONTRAST KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THE UP ARROW AND DOWN ARROW KEYPAD BUTTONS RESPONDED APPROPRIATELY TO BUTTONS PRESSES. THE KEYPAD COVER WAS INTACT; NO DAMAGE OR PEELING WAS OBSERVED. THE KEYPAD COVER WAS REMOVED AND CONTAMINATION WAS FOUND UNDER ALL KEY CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THE DISTRIBUTOR ALLEGED THAT THE DISPLAY SCREEN WAS DISCOLORED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
430337 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1