FDA Adverse Event Malfunction Summary report: N

AFFINITY BED

MDR report key: 3952226 · Received May 13, 2014

Report

Report Number
1824206-2014-01479
Event Type
Malfunction
Date Received
May 13, 2014
Date of Event
April 15, 2014
Report Date
April 15, 2014
Manufacturer
HILL-ROM, INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE BRAKE DETENT ASSEMBLY WAS BROKEN. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2009-2011 AND 2013 TO 2014. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE BRAKE DETENT ASSEMBLY TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECHNICIAN STATING THE BRAKES WERE NOT HOLDING. THE BED WAS LOCATED IN THE LABOR AND DELIVERY DEPARTMENT. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
285599 AFFINITY BED BIRTHING BED HDD HILL-ROM, INC. 3600

Patients

Seq Age Sex Outcome Treatment
1