FDA Adverse Event Malfunction Summary report: N

TOTALCARE BED

MDR report key: 3952207 · Received May 13, 2014

Report

Report Number
1824206-2014-01482
Event Type
Malfunction
Date Received
May 13, 2014
Date of Event
April 16, 2014
Report Date
April 16, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
PMA / PMN Number
K962942
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE COMMUNICATION CABLE AND COMMUNICATION BOARD WERE DAMAGED. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2012-2014. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE COMMUNICATION CABLE AND COMMUNICATION BOARD TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE NURSE CALL WOULD NOT WORK. THE BED WAS LOCATED IN THE ICU. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
285518 TOTALCARE BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1