FDA Adverse Event Malfunction Summary report: N

TRANSTAR STRETCHER

MDR report key: 3952199 · Received May 13, 2014

Report

Report Number
3006697241-2014-00443
Event Type
Malfunction
Date Received
May 13, 2014
Date of Event
April 15, 2014
Report Date
April 15, 2014
Manufacturer
HILLROM DE MEXICO S DE RL DE CV
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE BRAKE PADS WERE OUT OF ADJUSTMENT. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN ADJUSTED THE BRAKE PADS TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECHNICIAN STATING THE BRAKES WERE NOT HOLDING. THE BED WAS LOCATED ON THE 1ST FLOOR AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
286530 TRANSTAR STRETCHER WHEELED STRETCHER FPO HILLROM DE MEXICO S DE RL DE CV 8000

Patients

Seq Age Sex Outcome Treatment
1