FDA Adverse Event Malfunction Summary report: N

TOTALCARE BARIATRIC PLUS W/AIR BED

MDR report key: 3952183 · Received May 13, 2014

Report

Report Number
1824206-2014-01474
Event Type
Malfunction
Date Received
May 13, 2014
Date of Event
April 14, 2014
Report Date
April 14, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
PMA / PMN Number
K122473
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE POWER CORD HAD FLASH MARKS AND BURN MARKS ON THE PLUG AND ALSO THE POWER CORD PRONGS WERE BENT OVER, MOST LIKELY CAUSED BY DAMAGE OR USER ERROR SUCH AS THE BED NOT BEING UNPLUGGED BEFORE TRANSPORT. THE BED WAS SWAPPED FOR A KNOWN WORKING BED. THE TECHNICIAN REPLACED THE POWER CORD TO RESOLVE THE ISSUE AT THE HILL-ROM WAREHOUSE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

THE ACCOUNT REPORTED THAT SPARKS ARE COMING OUT OF THE PLUG ON THE POWER CORD. THE BED WAS LOCATED IN THE STORE ROOM ON THE FIRST FLOOR AT THE FACILITY. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
286861 TOTALCARE BARIATRIC PLUS W/AIR BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1840

Patients

Seq Age Sex Outcome Treatment
1