FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE BARIATRIC PLUS W/AIR BED
MDR report key: 3952183
·
Received May 13, 2014
Report
- Report Number
- 1824206-2014-01474
- Event Type
- Malfunction
- Date Received
- May 13, 2014
- Date of Event
- April 14, 2014
- Report Date
- April 14, 2014
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- PMA / PMN Number
- K122473
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND THE POWER CORD HAD FLASH MARKS AND BURN MARKS ON THE PLUG AND ALSO THE POWER CORD PRONGS WERE BENT OVER, MOST LIKELY CAUSED BY DAMAGE OR USER ERROR SUCH AS THE BED NOT BEING UNPLUGGED BEFORE TRANSPORT. THE BED WAS SWAPPED FOR A KNOWN WORKING BED. THE TECHNICIAN REPLACED THE POWER CORD TO RESOLVE THE ISSUE AT THE HILL-ROM WAREHOUSE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem · 1
THE ACCOUNT REPORTED THAT SPARKS ARE COMING OUT OF THE PLUG ON THE POWER CORD. THE BED WAS LOCATED IN THE STORE ROOM ON THE FIRST FLOOR AT THE FACILITY. THERE WAS NO PATIENT/USER INJURY REPORTED. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 286861 | TOTALCARE BARIATRIC PLUS W/AIR BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |