FDA Adverse Event Malfunction Summary report: N

PRIME CARE (R) B900

MDR report key: 3952146 · Received May 12, 2014

Report

Report Number
3007538326-2014-00041
Event Type
Malfunction
Date Received
May 12, 2014
Date of Event
April 23, 2014
Report Date
May 12, 2014
Manufacturer
PRIMUS MEDICAL, LLC
Product Code
FNL
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BED WAS INSPECTED BY PRIMUS MEDICAL ON (B)(4) 2014. THE INSPECTION DETERMINED THAT THE BRACKET WHERE THE FOOT HIGH-LOW ACTUATOR MOUNTS ONTO THE BACKBONE OF THE BED FRAME BROKE OFF. A NEW BED FRAME WAS DELIVERED TO THE FACILITY ON (B)(4) 2014. THIS PROBLEM HAS BEEN ASSIGNED CAPA (B)(4), AND A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF THE CORRECTIVE ACTION.

Description of Event or Problem · 1

CUSTOMER CALLED IN SAYING THAT THEY HAD A BED WITH A BROKEN WELD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
284591 PRIME CARE (R) B900 FNL PRIMUS MEDICAL, LLC PCB900

Patients

Seq Age Sex Outcome Treatment
1