FDA Adverse Event Injury Summary report: N

MIDMARK

MDR report key: 3952023 · Received July 16, 2014

Report

Report Number
1523530-2014-00012
Event Type
Injury
Date Received
July 16, 2014
Date of Event
June 16, 2014
Report Date
July 15, 2014
Manufacturer
MIDMARK CORP.
Product Code
NRU
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

AN EVAL OF THE CHAIR WAS DONE ON-SITE AND A REVIEW WAS CONDUCTED AT OUR FACILITY OF A COMPARABLE LOT. THE CHAIR DESIGN WAS REVIEWED FOR COMPLIANCE TO APPLICABLE SAFETY STANDARDS AND WAS FOUND TO BE IN COMPLIANCE. THE RISK MANAGEMENT FILE WAS REVIEWED AND FOUND TO HAVE ISSUE TO BE EVALUATED. THE FAILURE THAT OCCURRED HAS BEEN DETERMINED TO BE RELATED TO FAILURE TO FOLLOW USER MANUAL GUIDANCE.

Description of Event or Problem · 1

A DENTIST REPORTED THAT A PARENT WAS SEATED SIDEWAYS ON THE CHAIR WITH CHILD, WHO WAS THE PT, ON HIS LAP WHEN THE MIDMARK ELEVANCE DENTAL CHAIR WAS COMMENDED TO MOVE. THE PARENT'S LEG WAS PINCHED IN THE GAP BETWEEN THE CHAIR SEAT AND THE BACK. NO MEDICAL INTERVENTION WAS REQUIRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
415899 MIDMARK ELEVANCE DENTAL CHAIR NRU MIDMARK CORP. 153810

Patients

Seq Age Sex Outcome Treatment
1