ANIMAS VIBE
Report
- Report Number
- 2531779-2014-21054
- Event Type
- Malfunction
- Date Received
- July 23, 2014
- Report Date
- July 3, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS AT THE TIME OF THIS REPORT. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/11/2014 WITH THE FOLLOWING FINDINGS: MOISTURE CORROSION WAS OBSERVED ON THE DISPLAY SCREEN INSIDE THE PUMP. THE PUMP FAILED A LEAK TEST DUE A DISPLAY LENS LEAK. THE KEYPAD RUBBER WAS UNDAMAGED. THE PUMP ALARMED WITH A HARD CALL SERVICE ALARM. THE RESPONSIVENESS OF THE KEYPADS COULD NOT BE INVESTIGATED AS A RESULT OF THE CALL SERVICE ALARM. THE KEYPAD WAS REMOVED AND NO CONTAMINATION OR DAMAGE WAS FOUND TO THE KEY¿S CONTACTS. MOISTURE CORROSION WAS FOUND TO THE PRINTED CIRCUIT BOARD ON THE DISPLAY SCREEN, THE KEYPAD FLEX/CONNECTOR, THE CONTINUOUS GLUCOSE MONITOR MODULE, AND OTHER INTERNAL COMPONENTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2014, A REPORTER CONTACTED ANIMAS ALLEGING THAT THE DOWN, UP, AND OK BUTTONS WERE UNRESPONSIVE. THE REPORTER ADDITIONALLY ALLEGED THAT THERE WAS MOISTURE IN THE PUMP BEHIND THE DISPLAY SCREEN. THIS COMPLAINT IS BEING REPORTED BECAUSE IT WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO ALLEGATION OF A SERIOUS INJURY ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 431012 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 37 YR |