5MM MONOPOLAR HANDLE 33CM
Report
- Report Number
- 0002936485-2014-00527
- Event Type
- Malfunction
- Date Received
- July 23, 2014
- Date of Event
- July 7, 2014
- Report Date
- July 7, 2014
- Manufacturer
- STRYKER ENDOSCOPY-SAN JOSE
- Product Code
- KOG
- PMA / PMN Number
- K973259
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.
THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED. DURING THE INSPECTION OF THE 5MM MONOPOLAR HANDLE 33CM IT WAS CONFIRMED, THE INSULATION IS COMPROMISED. VISIBLE DINGS AND SCRATCHES WERE SEEN THROUGHOUT THE SHAFT. THE 5MM MONOPOLAR HANDLE 33CM FAILED THE INSULATION INTEGRITY TEST. IFU 1000401070 STATES "BEFORE USE, ENSURE THAT THERE ARE NO BREAKS, CHIPS, CRACKS, SCRATCHES, TEARS, OR MISSING/LOOSE COMPONENTS ON THE SHAFT INSULATION, HANDLE, OR HOUSING. DO NOT USE THE INSTRUMENT IF ANY OF THESE DEFECTS ARE PRESENT AS THIS COULD CAUSE UNINTENDED ELECTROSURGICAL BURNS AND LIFE THREATENING COMPLICATIONS. IF DAMAGE HAS OCCURRED, DISCONTINUE USE AND RETURN THE INSTRUMENT FOR REPAIR OR REPLACEMENT." THE PROBABLE ROOT CAUSE COULD BE WEAR AND TEAR AND OR MISHANDLING. IN SUM, THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED.
IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.
IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 431720 | 5MM MONOPOLAR HANDLE 33CM | ENDOSCOPE AND/OR ACCESSORIES | KOG | STRYKER ENDOSCOPY-SAN JOSE | 0643130D |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |