FDA Adverse Event Injury Summary report: N

6IN AML SM STATURE 12.0MM

MDR report key: 3951524 · Received July 23, 2014

Report

Report Number
1818910-2014-24016
Event Type
Injury
Date Received
July 23, 2014
Date of Event
April 4, 2013
Report Date
July 9, 2014
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
LPH
PMA / PMN Number
K061833
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A WORLDWIDE COMPLAINT DATABASE SEARCH FOUND NO OTHER REPORTED INCIDENT(S) AGAINST THE PROVIDED PRODUCT/LOT COMBINATION(S) SINCE RELEASE FOR DISTRIBUTION. BASED ON PREVIOUS INVESTIGATIONS THIS COMPLICATION OF JOINT REPLACEMENT IS UNLIKELY TO HAVE BEEN THE RESULT OF A DEVICE FAILING TO MEET REQUIRED SPECIFICATIONS. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. CORRECTIVE ACTION WAS NOT INDICATED. MEDICAL RECORDS WERE OBTAINED AND REVIEWED BY A MEDICAL PROFESSIONAL. FROM A MEDICAL PERSPECTIVE, BASED ON THE INFORMATION AVAILABLE, IT IS NOT POSSIBLE TO DETERMINE IF THE COMPLAINT WAS PRODUCT RELATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Description of Event or Problem · 1

PATIENT WAS REVISED TO ADDRESS PAIN AND INSTABILITY. UPDATE REC¿D (B)(4) 2014 - LITIGATION PAPERS RECEIVED. IN ADDITION TO WHAT WAS PREVIOUSLY REPORTED, LITIGATION ALLEGES THE PATIENT SUFFERS FROM A CLICKING NOISE, SLIPPING AND POPPING SENSATIONS, A GRATING AND GRINDING SENSATION AND ELEVATED METAL ION LEVELS. WE ARE MAKING THE HEAD AND LINER REPORTABLE TO ADDRESS THE METAL ON METAL ALLEGATIONS. THE COMPLAINT WAS UPDATED ON: (B)(4) 2014. UPDATE REC'D (B)(4) 2014 - PFS AND MEDICAL RECORDS RECEIVED. A CORRECT DOR WAS GIVEN. AFTER REVIEW OF THE MEDICAL RECORDS THE REVISION OPERATIVE NOTE INDICATED METALLOSIS AND CORROSION ON THE TRUNNION. THE STEM IS NOW BEING ADDED TO THE COMPLAINT. THERE IS NO NEW ADDITIONAL INFORMATION THAT WOULD AFFECT THE EXISTING MDR DECISION. THE COMPLAINT WAS UPDATED ON: (B)(4) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
431458 6IN AML SM STATURE 12.0MM HIP FEMORAL STEM/SLEEVE LPH DEPUY ORTHOPAEDICS INC US D3RE71000

Patients

Seq Age Sex Outcome Treatment
1 72 YR