6IN AML SM STATURE 12.0MM
Report
- Report Number
- 1818910-2014-24016
- Event Type
- Injury
- Date Received
- July 23, 2014
- Date of Event
- April 4, 2013
- Report Date
- July 9, 2014
- Manufacturer
- DEPUY ORTHOPAEDICS INC US
- Product Code
- LPH
- PMA / PMN Number
- K061833
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- PHYSICIAN
Narratives
ADDITIONAL NARRATIVE: IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A WORLDWIDE COMPLAINT DATABASE SEARCH FOUND NO OTHER REPORTED INCIDENT(S) AGAINST THE PROVIDED PRODUCT/LOT COMBINATION(S) SINCE RELEASE FOR DISTRIBUTION. BASED ON PREVIOUS INVESTIGATIONS THIS COMPLICATION OF JOINT REPLACEMENT IS UNLIKELY TO HAVE BEEN THE RESULT OF A DEVICE FAILING TO MEET REQUIRED SPECIFICATIONS. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. CORRECTIVE ACTION WAS NOT INDICATED. MEDICAL RECORDS WERE OBTAINED AND REVIEWED BY A MEDICAL PROFESSIONAL. FROM A MEDICAL PERSPECTIVE, BASED ON THE INFORMATION AVAILABLE, IT IS NOT POSSIBLE TO DETERMINE IF THE COMPLAINT WAS PRODUCT RELATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
PATIENT WAS REVISED TO ADDRESS PAIN AND INSTABILITY. UPDATE REC¿D (B)(4) 2014 - LITIGATION PAPERS RECEIVED. IN ADDITION TO WHAT WAS PREVIOUSLY REPORTED, LITIGATION ALLEGES THE PATIENT SUFFERS FROM A CLICKING NOISE, SLIPPING AND POPPING SENSATIONS, A GRATING AND GRINDING SENSATION AND ELEVATED METAL ION LEVELS. WE ARE MAKING THE HEAD AND LINER REPORTABLE TO ADDRESS THE METAL ON METAL ALLEGATIONS. THE COMPLAINT WAS UPDATED ON: (B)(4) 2014. UPDATE REC'D (B)(4) 2014 - PFS AND MEDICAL RECORDS RECEIVED. A CORRECT DOR WAS GIVEN. AFTER REVIEW OF THE MEDICAL RECORDS THE REVISION OPERATIVE NOTE INDICATED METALLOSIS AND CORROSION ON THE TRUNNION. THE STEM IS NOW BEING ADDED TO THE COMPLAINT. THERE IS NO NEW ADDITIONAL INFORMATION THAT WOULD AFFECT THE EXISTING MDR DECISION. THE COMPLAINT WAS UPDATED ON: (B)(4) 2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 431458 | 6IN AML SM STATURE 12.0MM | HIP FEMORAL STEM/SLEEVE | LPH | DEPUY ORTHOPAEDICS INC US | D3RE71000 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR |