NEOPUFF INFANT RESUSCITATOR
Report
- Report Number
- 9611451-2014-00645
- Event Type
- Malfunction
- Date Received
- July 22, 2014
- Report Date
- June 25, 2014
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTL
- PMA / PMN Number
- K892885
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
(B)(4). METHOD: THE COMPLAINT RD900 INFANT NEOPUFF RESUSCITATOR WAS RETURNED TO OUR SERVICE CENTER IN (B)(6) WHERE IT WAS VISUALLY INSPECTED BY A TRAINED FPH SERVICE ENGINEER. THEREFORE, OUR INVESTIGATION IS BASED ON THE INFORMATION PROVIDED BY OUR SERVICE CENTER. RESULTS: VISUAL INSPECTION REVEALED THAT THE GAS OUTLET PORT ON THE RD900 WAS BROKEN. THE UPPER END CAP WAS ALSO FOUND TO BE CRACKED. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT 120711. CONCLUSION: THE NEOPUFF IS A PORTABLE, REUSABLE DEVICE USED TO ASSIST IN THE DELIVERY OF RESPIRATORY BREATHS TO INFANTS UNTIL ADEQUATE SPONTANEOUS BREATHING OCCURS. BEING A PORTABLE DEVICE, THE NEOPUFF CAN BE SUSCEPTIBLE TO IMPACT DAMAGE, FOR INSTANCE WHEN DROPPED OR SUBJECTED TO CONSIDERABLE EXTERNAL FORCE. THE NEOPUFF TECHNICAL MANUAL STATES THE FOLLOWING: "DROPPING THE NEOPUFF INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS AS OUTLINED [IN THE MANUAL] BEFORE CONNECTION TO A PATIENT." ALL NEOPUFF UNITS ARE VISUALLY INSPECTED AND PERFORMANCE TESTED BEFORE LEAVING THE PRODUCTION LINE AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS THAT THE DAMAGE OCCURRED AFTER THE SUBJECT NEOPUFF UNIT WAS RELEASED FOR DISTRIBUTION. THE RETURNED NEOPUFF DEVICE WAS FITTED WITH A NEW FASCIA, VALVE SYSTEM, AND UPPER END CAP AND WAS RETURNED TO THE CUSTOMER AFTER PASSING THE SAFETY AND PERFORMANCE CHECKS AS PER THE NEOPUFF TECHNICAL MANUAL.
A HOSPITAL IN (B)(6) REPORTED THAT THE GAS OUTLET PORT OF AN RD900 NEOPUFF INFANT RESUSCITATOR WAS DAMAGED. A SERVICE OF THE DEVICE WAS REQUESTED.NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 428487 | NEOPUFF INFANT RESUSCITATOR | BTL | BTL | FISHER & PAYKEL HEALTHCARE LTD | RD900 | 120711 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |