MEGA NEEDLE DRIVER INSTRUMENT
Report
- Report Number
- 2955842-2014-04490
- Event Type
- Malfunction
- Date Received
- July 22, 2014
- Date of Event
- June 26, 2014
- Report Date
- June 26, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS FOUND THAT ONE GRIP CLOSE CABLE WAS BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY SPUN FREELY AND DID NOT EXHIBIT DAMAGE. THE CABLE SEGMENT STUCK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST WERE UNDAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE DAMAGED CABLE COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.
IT WAS REPORTED THAT THE MEGA NEEDLE DRIVER INSTRUMENT WIRE WAS STICKING OUT OF THE INSTRUMENT DURING A DA VINCI SURGICAL PROCEDURE. NO FRAGMENTS FELL INTO THE PATIENT. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 428757 | MEGA NEEDLE DRIVER INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420194-10 | M10140307 617 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |