FDA Adverse Event Malfunction Summary report: N

VENTED AUTOFEED CHAMBER

MDR report key: 3951468 · Received July 22, 2014

Report

Report Number
9611451-2014-00644
Event Type
Malfunction
Date Received
July 22, 2014
Date of Event
June 23, 2014
Report Date
June 24, 2014
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT MR290V VENTED HUMIDIFICATION CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION. IT WAS VISUALLY INSPECTED AND SUBSEQUENTLY CONNECTED TO A WATER BAG TO TEST FOR LEAK. RESULTS: VISUAL INSPECTION REVEALED TWO HOLES IN THE FILTER OF THE BAG SPIKE VENT. UPON CONNECTING TO A WATER BAG, WATER DROPLETS BEGAN TO FORM AT THE SPIKE VENT. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT 140204. CONCLUSION: WE ARE UNABLE TO DETERMINE THE CAUSE OF THE DAMAGE OBSERVED ON THE RETURNED BAG SPIKE VENT. ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. ANY CHAMBER THAT FAILS IS REJECTED. THIS SUGGESTS THAT THE DAMAGE TO THE BAG SPIKE VENT OCCURRED AFTER THE CHAMBER WAS RELEASED FOR DISTRIBUTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE MR290 STATE THE FOLLOWING: -"SET APPROPRIATE VENTILATOR ALARMS." -"PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." THE HOSPITAL STAFF CORRECTLY CHECKED THE MR290 PRIOR TO PATIENT USE. THIS IS IN LINE WITH OUR USER INSTRUCTIONS.

Description of Event or Problem · 1

A HEALTHCARE FACILITY IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) REPRESENTATIVE THAT WATER LEAKED FROM THE VENT VALVE OF THE BAG SPIKE ON AN MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER.THIS WAS FOUND PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
428513 VENTED AUTOFEED CHAMBER AUTOFEED CHAMBER BTT FISHER & PAYKEL HEALTHCARE LTD MR290V 1402040304

Patients

Seq Age Sex Outcome Treatment
1