PROGRASP FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2014-04487
- Event Type
- Malfunction
- Date Received
- July 22, 2014
- Date of Event
- June 24, 2014
- Report Date
- June 24, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS FOUND THAT THE PITCH DOWN CABLE WAS FRAYED AT THE DISTAL CLEVIS HUB. THE FRAYED STRANDS STUCK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST WERE UNDAMAGED. FAILURE ANALYSIS ALSO OBSERVED THAT THE DISTAL END OF THE MAIN TUBE HAD VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE .071 - .977 IN LENGTH AND NOT ALIGNED WITH THE TUBE AXIS. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE DAMAGED CABLE AND/OR THE TUBE ABRASIONS, FOUND DURING FAILURE ANALYSIS EVALUATION COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.
IT WAS REPORTED THAT THE PROGRASP FORCEPS INSTRUMENT CABLE WAS NOTED TO BE BROKEN DURING A DA VINCI SURGICAL PROCEDURE. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO PATIENT OR PATIENT HARM. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 428967 | PROGRASP FORCEPS INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420093-10 | M11121026 256 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |