FDA Adverse Event Malfunction Summary report: N

RESERVOIR 1.8ML

MDR report key: 3951363 · Received July 22, 2014

Report

Report Number
3004209178-2014-88088
Event Type
Malfunction
Date Received
July 22, 2014
Date of Event
June 24, 2014
Report Date
June 24, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K001828
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

INSPECTED RESERVOIR, SNAP-CAP AND SEPTUM FOR ANOMALIES NONE WERE FOUND. PERFORMED OCCLUSION TEST PER SPECIFICATIONS, RESERVOIR FILLED AND CONNECTED TO NEW INFUSION SET. INSTALLED RESERVOIR AND CONNECTOR INTO INSULIN PUMP AND PERFORMED INFUSION SET. INSTALLED RESERVOIR AND CONNECTOR TO INSULIN PUMP AND PERFORMED CATHETER TIP. RESERVOIR NOT OCCLUDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN LEAKED INTO THE RESERVOIR COMPARTMENT FROM THE RESERVOIR BARREL. THE CUSTOMER'S BLOOD GLUCOSE WAS 260 MG/DL. THE CUSTOMER WAS ADVISED TO DISCONNECT FROM THE INSULIN PUMP, TO REMOVE THE RESERVOIR AND TO DRY THE RESERVOIR COMPARTMENT WITH A LINT FREE CLOTH OR STERILE GAUZE. THE CUSTOMER STATED THAT THEY USED NAPKIN AND THERE WERE NO DEBRIS LEFT OVER. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
428447 RESERVOIR 1.8ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-326A H8489386

Patients

Seq Age Sex Outcome Treatment
1 39 YR