FDA Adverse Event
Malfunction
Summary report: N
ENDOTAK RELIANCE
MDR report key: 3951205
·
Received July 22, 2014
Report
- Report Number
- 2124215-2014-14254
- Event Type
- Malfunction
- Date Received
- July 22, 2014
- Date of Event
- June 12, 2014
- Report Date
- June 13, 2014
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD AND IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED HIGH SHOCKING LEAD IMPEDANCE MEASUREMENTS GREATER THAN 125 OHMS. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) PROVIDED TROUBLESHOOTING OPTIONS. ADDITIONAL INFORMATION INDICATED THAT NO INTERVENTION WAS PERFORMED AT THIS TIME AND THE PATIENT WILL CONTINUE TO BE MONITORED. THE SYSTEM REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 428375 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | LWS | CPI - DEL CARIBE | 0158 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | 1488T| 0158| T165| E163| A135 |