HOMECHOICE PRO
Report
- Report Number
- 1416980-2014-23719
- Event Type
- Malfunction
- Date Received
- July 22, 2014
- Date of Event
- May 12, 2014
- Report Date
- May 12, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NM, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). ADDITIONAL INFORMATION: THE HOMECHOICE (HC) DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB (PAL). THE EVENT HISTORY LOG REVIEW SHOWED NO KEYSTROKES, PROGRAMMING, OR USE RELATED EVENTS THAT INDICATED AND/OR CONTRIBUTED TO THE REPORTED ISSUE. THE EXTERNAL/INTERNAL INSPECTION WAS PERFORMED AND PASSED, ALONG WITH ALL OF THE ELECTRICAL TESTING. THE FLUID VOLUMETRIC ACCURACY TEST WAS PERFORMED AND THE DEVICE FAILED; THEREFORE, PAL PERFORMED A MORE DETAILED INSPECTION OF THE DOOR ASSEMBLY. THE INSPECTION SHOWED THAT THERE WAS AN INSUFFICIENT AMOUNT OF COPPER MESH. THE DEVICE COULD NOT DELIVER AN ACCURATE AMOUNT OF FLUID DUE TO THE INSUFFICIENT AMOUNT OF COPPER MESH RESULTING IN THE VOLUMETRIC ACCURACY FAILURE. THE DOOR PISTON WAS TO BE SCRAPPED AND THE DEVICE WAS SENT FOR SERVICING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
(B)(4). THE HOMECHOICE (HC) WAS RETURNED AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE (HC) DEVICE, A BAXTER TECHNICIAN DETERMINED THE HC MACHINE FAILED THE VOLUME ACCURACY TEST. THE DEVICE FAILED DURING EVALUATION WITH NO PATIENT INVOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 427674 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |