HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2014-23715
- Event Type
- Malfunction
- Date Received
- July 22, 2014
- Date of Event
- June 28, 2014
- Report Date
- June 29, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. THE DEVICE WAS NOT RETURNED; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. HOWEVER, IMPROPERLY DISCONNECTING/RECONNECTING FROM THE SET IS A KNOWN CAUSE OF THIS ALARM. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿ WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY DISCONNECTING DURING EMERGENCIES AND FOR RETURNING TO THERAPY AFTER THE EMERGENCY DISCONNECT PROCEDURE. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A SYSTEM ERROR 2240 (AIR IN SET/LINE) ALARM OCCURRED ON THE HOMECHOICE (HC) DEVICE DURING DWELL 1. THE HOME PATIENT (HP) WAS CONNECTED AT THE TIME OF THE ALARM. DURING TROUBLESHOOTING, IT WAS NOTED THAT THE PATIENT DID NOT USE ASEPTIC TECHNIQUE WHEN DISCONNECTING AND THEN RECONNECTED WITHOUT USING ASEPTIC TECHNIQUE. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) REVIEWED PROPER CONNECTION/DISCONNECTION PROCEDURES WITH THE REPORTER. THE TSR HAD THE HP CYCLE POWER ON THE HC DEVICE, CLOSE ALL CLAMPS, AND ASSISTED THE HP IN SAFELY REMOVING THE CASSETTE TO CLEAR THE ALARM. THE HP WOULD COMPLETE THERAPY USING ALL NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 428357 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | HOMECHOICE |