HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2014-23714
- Event Type
- Malfunction
- Date Received
- July 22, 2014
- Date of Event
- June 28, 2014
- Report Date
- June 28, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP MDR WILL BE SUBMITTED.
(B)(4). THE SAMPLE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN; THEREFORE, A DEVICE EVALUATION COULD NOT BE PERFORMED. HOWEVER, A LOOSE CONNECTION IS A KNOWN CAUSE OF THE REPORTED ALARM. THE CAUSE OF LOOSE CONNECTION COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW UP REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A HOMECHOICE DEVICE EXPERIENCED A SYSTEM ERROR 2240 (AIR IN LINE/SET). THE PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. THIS OCCURRED DURING THE INITIAL DRAIN OF PERITONEAL DIALYSIS THERAPY. DURING TROUBLESHOOTING IT WAS FOUND THERE WAS A LOOSE CONNECTION. THE PATIENT WAS TO START OVER WITH NEW SUPPLIES. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION OR ADVERSE EVENT ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 427956 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | HOMECHOICE |