FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3951104 · Received July 22, 2014

Report

Report Number
1416980-2014-23714
Event Type
Malfunction
Date Received
July 22, 2014
Date of Event
June 28, 2014
Report Date
June 28, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP MDR WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN; THEREFORE, A DEVICE EVALUATION COULD NOT BE PERFORMED. HOWEVER, A LOOSE CONNECTION IS A KNOWN CAUSE OF THE REPORTED ALARM. THE CAUSE OF LOOSE CONNECTION COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A HOMECHOICE DEVICE EXPERIENCED A SYSTEM ERROR 2240 (AIR IN LINE/SET). THE PATIENT WAS CONNECTED AT THE TIME OF THE ALARM. THIS OCCURRED DURING THE INITIAL DRAIN OF PERITONEAL DIALYSIS THERAPY. DURING TROUBLESHOOTING IT WAS FOUND THERE WAS A LOOSE CONNECTION. THE PATIENT WAS TO START OVER WITH NEW SUPPLIES. THERE WAS NO PATIENT INJURY, MEDICAL INTERVENTION OR ADVERSE EVENT ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
427956 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 66 YR HOMECHOICE