FLOGARD
Report
- Report Number
- 1416980-2014-23716
- Event Type
- Malfunction
- Date Received
- July 22, 2014
- Date of Event
- June 28, 2014
- Report Date
- June 28, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS SERVICED ON-SITE BY A FIELD SERVICE TECHNICIAN. VISUAL INSPECTION, FUNCTIONAL TESTING, AND A REVIEW OF THE ALARM LOG WERE PERFORMED. FUNCTIONAL TESTING REVEALED THAT THE DEVICE HAD A F-38 ALARM. THE CAUSE OF THE CONDITION WAS DETERMINED TO BE A DAMAGED FORCE SENSING RESISTOR. TO CORRECT THE CONDITION, THE FORCE SENSING RESISTOR WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A FLOGARD INFUSION PUMP HAD A F-38 ALARM. IT WAS NOT SPECIFIED WHEN IN THE PROCESS THIS OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION IS AVAILABLE.