MELODY TRANSCATHER PULMONARY

MDR report key: 3951079 · Received July 22, 2014

Report

Report Number
2025587-2014-00504
Event Type
Injury
Date Received
July 22, 2014
Date of Event
April 4, 2012
Report Date
August 13, 2014
Manufacturer
MEDTRONIC HEART VALVES DIVISION
Product Code
NPV
PMA / PMN Number
H080002
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AFTER EXPLANT, THE TPV WAS FORWARDED TO ANOTHER FACILITY FOR ANALYSIS. A SUPPLEMENTAL REPORT WILL BE FILED WHEN THE INVESTIGATION IS COMPLETED OR IF THE DEVICE IS RETURNED TO MEDTRONIC FOR ANALYSIS. (B)(4).

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

HAEMOPHILUS PARAINFLUENZAE IS PART OF THE NORMAL HUMAN OROPHARYNX FLORA AND IS A MEMBER OF THE HACEK GROUP OF RARE BACTERIA THAT IS RESPONSIBLE FOR ABOUT 3% OF ENDOCARDITIS CASES. THE DEVICE HISTORY RECORD WAS REVIEWED AND SHOWED THAT THIS PRODUCT MET ALL MANUFACTURING SPECIFICATIONS FOR PRODUCT RELEASED FOR DISTRIBUTION. NO ISSUES WERE IDENTIFIED THAT WOULD HAVE IMPACTED THIS EVENT. A REVIEW OF THE STERILITY LOT RECORD CONFIRMED THAT THE BIOLOGICAL INDICATORS (BI) TESTED NEGATIVE FOR BOTH TISSUE AND SOLUTION. THERE HAVE BEEN NO ADDITIONAL COMPLAINTS, OR COMPLAINTS FOR INFECTION AGAINST ANY OF THE DEVICES MANUFACTURED UNDER THIS STERILITY LOT. THE STERILIZATION PROCESS USED BY MEDTRONIC HAS AN ANTI-MICROBIAL KILL ON THE MICROORGANISMS SUCH AS STAPHYLOCOCCUS AND STREPTOCOCCUS SPECIES. THE INFORMATION RECEIVED INDICATES THAT THE ENDOCARDITIS OCCURRED 2 YEARS POST IMPLANT. ENDOCARDITIS THAT OCCURS MORE THAN 12 MONTH AFTER THE PROCEDURE ARE CALLED LATE PROSTHETIC-VALVE ENDOCARDITIS AND ARE LARGELY COMMUNITY-ACQUIRED VERSUS A RESULT OF THE MANUFACTURING PROCESS OF THE VALVE. BASED ON THE ABOVE INVESTIGATION, THE ENDOCARDITIS IS UNLIKELY TO BE ATTRIBUTED TO THE MELODY TPV DEVICE AND/OR MANUFACTURING PROCESS. A ROOT CAUSE OF THE REPORTED CLINICAL OBSERVATIONS COULD NOT BE DETERMINED. HOWEVER, THERE IS NO INDICATION THAT THE REPORTED ENDOCARDITIS WAS RELATED TO THE MANUFACTURE OF THE DEVICE. THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS/INVESTIGATION WILL BE UPDATED AND A SUPPLEMENTAL REPORT SUBMITTED IF THE DEVICE IS RETURNED.

Description of Event or Problem · 1

MEDTRONIC RECEIVED INFORMATION THAT THIS TRANSCATHETER PULMONARY VALVE (TPV) WAS IMPLANTED INTO A STENOTIC CONDUIT AFTER THE CONDUIT WAS PRE-STENTED. TWO YEARS POST-IMPLANT, THE PATIENT WAS ADMITTED TO A HOSPITAL WITH A FEVER. AN ECHOCARDIOGRAM SHOWED NO VEGETATION AND BLOOD CULTURES WERE STERILE. ANTIBIOTICS WERE PRESCRIBED WITH NO IMPROVEMENT. THE PATIENT WAS TRANSFERRED TO THE DEVICE IMPLANT CENTER, WHERE AN ECHOCARDIOGRAM SHOWED THE TPV WAS STENOTIC, AND HEMOPHILUS PARAINFLUENZA WAS IDENTIFIED IN BLOOD CULTURES. AFTER SUCCESSFUL TREATMENT WITH ANTIBIOTICS, THE TPV WAS EXPLANTED AND A NON-MEDTRONIC AORTIC HOMOGRAFT WAS IMPLANTED INSIDE THE CONDUIT. AT EXPLANT, THE TPV REVEALED PARTIALLY-DESTRUCTED THICKENED LEAFLETS WITH A YELLOW SMEAR. CULTURES OF THE SMEAR REMAINED STERILE. NO SUBSEQUENT ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
427933 MELODY TRANSCATHER PULMONARY PULMONARY VALVE PROSTHESIS PERCUTANEOUSLY DELIVERED NPV MEDTRONIC HEART VALVES DIVISION PB1018

Patients

Seq Age Sex Outcome Treatment
1 00025 YR Required Intervention