FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3951077 · Received July 22, 2014

Report

Report Number
2531779-2014-20998
Event Type
Malfunction
Date Received
July 22, 2014
Report Date
June 27, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED FOR INVESTIGATION AND DEVICE EVALUATION WAS COMPLETED BY PRODUCT ANALYSIS ON 07/15/2014 WITH THE FOLLOWING FINDINGS: UNEXPLAINED POWER ON RESETS OBSERVED IN THE BLACK BOX. NO DAMAGE FOUND TO THE BATTERY COMPARTMENT. RETURNED BATTERY CAP HAS STRIPPED THREADS AND IS UNABLE TO FULLY SECURE TO THE PUMP; POWER ON RESETS DUPLICATED WITH RETURNED BATTERY CAP. A NEW TEST BATTERY CAP WAS ABLE TO FULLY ATTACH TO THE PUMP WITH NO YELLOW O RING SHOWING. A 24HR DURATION TEST WAS COMPLETED USING THE NEW TEST BATTERY CAP; NO POWER INTERRUPTIONS WERE DUPLICATED. REMOVAL OF THE PUMP COVER SHOWED NO EVIDENCE OF INTERNAL MOISTURE. NO DAMAGE TO THE POWER CIRCUIT WAS OBSERVED.

Description of Event or Problem · 1

THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED FOR PRODUCT ANALYSIS INVESTIGATION AND THE EVALUATION REVEALED A DAMAGED BATTERY CAP. THIS REPORT IS MADE BASED ON THE RESULTS OF AN INVESTIGATION COMPLETED ON 07/15/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
428273 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 11 YR