FDA Adverse Event Malfunction Summary report: N

NTLC LINEAR CUTTER

MDR report key: 3951071 · Received July 22, 2014

Report

Report Number
3005075853-2014-05156
Event Type
Malfunction
Date Received
July 22, 2014
Date of Event
July 8, 2014
Report Date
July 14, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K092577
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: THE ANALYSIS RESULTS FOUND THAT THE NTLC75 DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION AND WITH A RELOAD LOADED IN THE DEVICE. THE RELOAD WAS RETURNED FULLY FIRED. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED ALL THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. A BATCH RECORD REVIEW WAS PERFORMED AND THE BATCH HAD NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). ATTEMPTS HAVE BEEN MADE TO RETRIEVE THE DEVICE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. AT THE TIME OF THIS SUBMISSION, THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN OPEN TOTAL GASTRECTOMY, STAPLES WERE OVERLY BENT AT THE 3RD FIRING ON THE DUODENUM. AS THE LEGS WERE PROTRUDED FROM THE FIRED TISSUE, BURIED SUTURE WAS PERFORMED. THE STAPLE HEIGHT WAS GREEN. THE TARGET TISSUE WAS ENLARGED. REINFORCEMENT MATERIAL WAS NOT USED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
428337 NTLC LINEAR CUTTER STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA L4EG1P

Patients

Seq Age Sex Outcome Treatment
1 RELOAD - SR75 (LOT#:L4EH88)X1