FDA Adverse Event
Injury
Summary report: N
CAPSUREEPI
MDR report key: 3950920
·
Received July 22, 2014
Report
- Report Number
- 2182208-2014-01859
- Event Type
- Injury
- Date Received
- July 22, 2014
- Date of Event
- June 26, 2014
- Report Date
- June 26, 2014
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- DTB
- PMA / PMN Number
- P950024
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE CLINIC WITH COMPLAINTS OF A STOMACH ACHE. UPON EXAMINATION IT WAS OBSERVED THAT THE PATIENT'S HEART RATE WAS IN THE 40'S. PACEMAKER TRANSMISSION REVEALED THE RIGHT VENTRICULAR (RV) LEAD DISPLAYED HIGH IMPEDANCE. A CHEST X-RAY REVEALED THE LEAD WAS FRACTURED. THE LEAD WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 428590 | CAPSUREEPI | ELECTRODE, PACEMAKER, PERMANENT | DTB | MEDTRONIC, INC. | 4968-25 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00006 YR | Hospitalization| R | ADDR01 IPG, 4968-25 LEAD |