FDA Adverse Event
Malfunction
Summary report: N
HUDSON CONCHA NEPTUNE
MDR report key: 3950632
·
Received June 5, 2014
Report
- Report Number
- 3003898360-2014-00386
- Event Type
- Malfunction
- Date Received
- June 5, 2014
- Date of Event
- May 16, 2014
- Report Date
- May 29, 2014
- Manufacturer
- TELEFLEX
- Product Code
- BTT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
A VISUAL, FUNCTIONAL AND DIMENSIONAL INSPECTION OF THE PRODUCT INVOLVED IN THE COMPLAINT COULD NOT BE CONDUCTED SINCE THE PRODUCT WAS NOT RETURNED. DHR (DEVICE HISTORY RECORD) REVIEW COULD NOT BE CONDUCTED SINCE THE LOT NUMBER WAS NOT PROVIDED. A DOCUMENT ASSESSMENT (FMEA) WAS CONDUCTED AND NO CHANGES WERE REQUIRED. THE COMPLAINT CAN NOT BE CONFIRMED SINCE THE SAMPLE WAS NOT AVAILABLE FOR INVESTIGATION, THEREFORE, IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE FOR THE DEFECT REPORTED AND A CORRECTIVE ACTION FOR IT. TELEFLEX WILL CONTINUE TO MONITOR AND TREND RELATING COMPLAINTS.
Description of Event or Problem · 1
THE EVENT IS REPORTED AS: THE CUSTOMER ALLEGES THAT DURING PRE-TESTING, THE DEVICE IS CUTTING OFF SPORADICALLY AFTER BEING TURNED ON.