FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3950397 · Received July 22, 2014

Report

Report Number
2531779-2014-20954
Event Type
Malfunction
Date Received
July 22, 2014
Report Date
July 11, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP¿S BLACK BOX HISTORY SHOWED THAT POWER REBOOT OCCURRED AFTER A REPLACE BATTERY ALARM ON (B)(6) 2014 AND ONE UNEXPLAINED POWER REBOOT ON (B)(6) 2014. DURING TESTING, THE PUMP POWERED ON TO THE ¿VERIFY¿ SCREEN WITH AUDIO TONES AND VIBRATIONS FUNCTIONAL. THE BATTERY COMPARTMENT WAS INTACT; NO DAMAGE WAS OBSERVED. THE RETURNED BATTERY CAP WAS UNDAMAGED AND ABLE TO BE FULLY SECURED TO THE PUMP. THE PUMP WAS EXERCISED FOR 24 HOURS ON A 1 UNIT PER HOUR BASAL RATE WITH NO POWER INTERRUPTIONS OCCURRING. THE PUMP CASE WAS REMOVED AND NO EVIDENCE OF INTERNAL MOISTURE OR DAMAGE TO THE POWER CIRCUIT WAS FOUND. THE COMPLAINT THAT THE PUMP HAD NO POWER WAS OBSERVED IN THE PUMP HISTORY BUT WAS NOT ABLE TO BE DUPLICATED DURING INVESTIGATION.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (NO POWER) ISSUE. THE REPORTER ALLEGED THAT THE PUMP HAD NO POWER. IT WAS NOTED THAT A NEW BATTERY WAS INSERTED AND THE PUMP POWERED ON APPROPRIATELY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
428325 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 26 YR