FDA Adverse Event Injury Summary report: N

HYDROVIEW INTRAOCULAR LENS

MDR report key: 3946371 · Received July 10, 2014

Report

Report Number
1119279-2014-00165
Event Type
Injury
Date Received
July 10, 2014
Report Date
June 12, 2014
Manufacturer
BAUSCH & LOMB, INC.
Product Code
HQL
PMA / PMN Number
P990014
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE LENS WAS IMPLANTED IN 2001. HYDROVIEW IOL WAS DISCONTINUED AND IS NO LONGER MANUFACTURED BY B+L. DESPITE ATTEMPTS TO OBTAIN THE LENS LOT NUMBER, IT WAS NOT PROVIDED. THEREFORE, A REVIEW OF THE LOT HISTORY RECORD COULD NOT BE PERFORMED. THE LENS IS NOT AVAILABLE TO BE RETURNED TO B+L FOR EVALUATION. THEREFORE, IT IS NOT POSSIBLE TO CONFIRM WHETHER THE REPORT OPACIFICATION IS DUE TO CALCIFICATION OR SOME OTHER MECHANISM. THE CAUSE OF THE CALCIFICATION PHENOMENON IS MULTI-FACTORIAL. THREE MAJOR TYPES OF CALCIFICATION HAVE BEEN PREVIOUSLY DESCRIBED: THE PRIMARY FORM REFERS TO CALCIFICATION DUE TO ISSUES INHERENT TO THE IOL. THE SECONDARY FORM REFERS TO DEPOSITION OF CALCIUM ONTO THE SURFACE OF THE IOL MOST LIKELY DUE TO ENVIRONMENTAL CIRCUMSTANCES (E.G. CHANGES IN THE AQUEOUS SURROUNDING THE IMPLANTED IOL ASSOCIATED WITH PRE-EXISTING OR CONCURRENT DISEASES). BY DEFINITION, IT IS NOT RELATED TO ANY PROBLEM WITH THE IOL ITSELF. THE THIRD FORM IS A FALSE-POSITIVE OR PSEUDO-CALCIFICATION REFERS TO THOSE CASES IN WHICH OTHER PATHOLOGY IS MISTAKEN FOR CALCIFICATION OR FALSE-POSITIVE STAINING FOR CALCIUM. BASED ON THE INFORMATION PROVIDED, THE EXACT ROOT CAUSE FOR THIS EVENT COULD NOT BE DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A HYDROVIEW INTRAOCULAR LENS IS SCHEDULED TO BE EXPLANTED DUE TO OPACIFICATION. THE LOT AND SERIAL NUMBER WERE NOT PROVIDED, THEREFORE, IT HAS NOT BEEN POSSIBLE TO DETERMINE WHETHER THE LENS MODEL IS HYDROVIEW 1.0. ADDITIONAL INFORMATION WAS REQUESTED, BUT HS NOT BEEN RECEIVED TO DATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
403857 HYDROVIEW INTRAOCULAR LENS HQL/INTRAOCULAR LENS HQL BAUSCH & LOMB, INC. H60M

Patients

Seq Age Sex Outcome Treatment
1 Other