FDA Adverse Event Malfunction Summary report: N

GREENLIGHT MOXY FIBER OPTIC

MDR report key: 3946249 · Received July 21, 2014

Report

Report Number
2937094-2014-00644
Event Type
Malfunction
Date Received
July 21, 2014
Date of Event
July 14, 2014
Report Date
July 14, 2014
Manufacturer
AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE)
Product Code
GEX
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

PRODUCT EVALUATION: FAILURE ANALYSIS FOR FIBER (B)(4): THE FIBER SHOWS A CIRCUMFERENTIAL FRACTURE DISTAL TO FIBER/GLASS CAP FUSION ZONE PROXIMAL TO THE BEVEL EDGE; THE FIBER PROXIMAL TO FRACTURE CAN ROTATE INDEPENDENTLY OF METAL CAP; THE DISTAL END OF THE FIBER/GLASS CAP IS DETACHED AND RETURNED; THE GLASS CAP EXHIBITS SEVERE DEVITRIFICATION AT THE OUTPUT WINDOW; THE METAL CAP EXHIBITS SEVERE CHAR AND MELTED METAL AT THE OUTPUT WINDOW; THE OUTER FLOW TUBING EXHIBITS MILD CONTAMINATION, LIKELY BIOLOGIC. BASED ON THE ANALYSIS ABOVE, THE POTENTIAL FOR FORWARD FIRING MAY EXIST. PROBABLE ROOT CAUSE: BASED ON THE PRODUCT ANALYSIS RESULTS, THE PROBABLE ROOT CAUSE OF THE FAILURE IS: HEAT ACCUMULATION. CAP WEAR WAS ACCELERATED DUE TO ANATOMICAL/PROCEDURAL FACTORS (TISSUE CONTACT AND TECHNIQUE) ENCOUNTERED DURING THE PROCEDURE WHICH WOULD LIMIT THE PERFORMANCE OF THE FIBER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INNER CAP ON BOTH THE FIRST AND SECOND FIBERS FELL OFF AND CAUSED THE FIBERS TO FORWARD FIRE. THE CASE WAS COMPLETED WITH AN ALTERNATE PROCEDURE (TURP). THERE WAS NO PATIENT INJURY REPORTED. THIS REPORT IS FOR THE FIRST FIBER USED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
424854 GREENLIGHT MOXY FIBER OPTIC POWERED LASER SURGICAL INSTRUMENT GEX AMERICAN MEDICAL SYSTEMS, INC. (SAN JOSE) 0010-2400 411B

Patients

Seq Age Sex Outcome Treatment
1